Selection
SCPI Command :
MEASurement<*>:EMC:LIMit:SELection
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(measurement=Measurement.Default) EmcLimitsStandard[source]
# MEASurement<*>:EMC:LIMit:SELection value: enums.EmcLimitsStandard = driver.measurement.emc.limit.selection.get(measurement = repcap.Measurement.Default)
Selects the electrostatic discharge (ESD) immunity test levels according to the guidelines of the EN IEC 61000-4-2:2025 standard. For contact discharge (CD) and air discharge (AD) , it specifies the maximum voltage (kV) .
CD1: Contact discharge, 2 kV.
CD2: Contact discharge, 4 kV.
CD3: Contact discharge, 6 kV.
CD4: Contact discharge, 8 kV.
AD1: Air discharge, 2 kV.
AD2: Air discharge, 4 kV.
AD3: Air discharge, 8 kV.
AD4: Air discharge, 15 kV.
- Parameters:
measurement – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Measurement’)
- Returns:
limits_standard: CD1 | CD2 | CD3 | CD4 | AD1 | AD2 | AD3 | AD4
- set(limits_standard: EmcLimitsStandard, measurement=Measurement.Default) None[source]
# MEASurement<*>:EMC:LIMit:SELection driver.measurement.emc.limit.selection.set(limits_standard = enums.EmcLimitsStandard.AD1, measurement = repcap.Measurement.Default)
Selects the electrostatic discharge (ESD) immunity test levels according to the guidelines of the EN IEC 61000-4-2:2025 standard. For contact discharge (CD) and air discharge (AD) , it specifies the maximum voltage (kV) .
CD1: Contact discharge, 2 kV.
CD2: Contact discharge, 4 kV.
CD3: Contact discharge, 6 kV.
CD4: Contact discharge, 8 kV.
AD1: Air discharge, 2 kV.
AD2: Air discharge, 4 kV.
AD3: Air discharge, 8 kV.
AD4: Air discharge, 15 kV.
- Parameters:
limits_standard – CD1 | CD2 | CD3 | CD4 | AD1 | AD2 | AD3 | AD4
measurement – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Measurement’)