Dpoperator

SCPI Command :

TRIGger<*>:I2C:DPOPerator
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(trigger=Trigger.Default) DataIndexOperator[source]
# TRIGger<*>:I2C:DPOPerator
value: enums.DataIndexOperator = driver.trigger.i2C.dpoperator.get(trigger = repcap.Trigger.Default)

Sets the operator for the data position. You can defined an exact position, or a position range.

Parameters:

trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)

set(dat_posi_optor: DataIndexOperator, trigger=Trigger.Default) None[source]
# TRIGger<*>:I2C:DPOPerator
driver.trigger.i2C.dpoperator.set(dat_posi_optor = enums.DataIndexOperator.ANY, trigger = repcap.Trigger.Default)

Sets the operator for the data position. You can defined an exact position, or a position range.

Parameters:
  • dat_posi_optor

    ANY | OFF | EQUal | GETHan | INRange | RANGe

    • ANY = OFF: The position of the required pattern is not relevant.

    • EQUal | GETHan: Equal, Greater or equal than. These conditions require one data position to be set with TRIGgert:I2C:DPOSition.

    • INRange = RANGe: In range: Set the minimum and maximum value of the range with TRIGgert:I2C:DPOSition and TRIGgert:I2C:DPTO.

  • trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)