Dmin
SCPI Command :
TRIGger<*>:I2C:DMIN
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(trigger=Trigger.Default) int[source]
# TRIGger<*>:I2C:DMIN value: int = driver.trigger.i2C.dmin.get(trigger = repcap.Trigger.Default)
Specifies the data bit pattern, or sets the start value of a data pattern range.. Enter the bytes in msb first bit order. The maximum pattern length is 64 bit. Waveform data is compared with the pattern byte-by-byte.
- Parameters:
trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)
- Returns:
data: Numeric or string pattern, see ‘Bit pattern parameter’. The string parameter accepts the bit value X (don’t care) .
- set(data: int, trigger=Trigger.Default) None[source]
# TRIGger<*>:I2C:DMIN driver.trigger.i2C.dmin.set(data = 1, trigger = repcap.Trigger.Default)
Specifies the data bit pattern, or sets the start value of a data pattern range.. Enter the bytes in msb first bit order. The maximum pattern length is 64 bit. Waveform data is compared with the pattern byte-by-byte.
- Parameters:
data – Numeric or string pattern, see ‘Bit pattern parameter’. The string parameter accepts the bit value X (don’t care) .
trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)