FilterPy

SCPI Command :

TRIGger<*>:ANEDge:FILTer
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(trigger=Trigger.Default) TriggerCoupling[source]
# TRIGger<*>:ANEDge:FILTer
value: enums.TriggerCoupling = driver.trigger.anEdge.filterPy.get(trigger = repcap.Trigger.Default)

Sets a filter for the external trigger signal to reject high or low frequencies.

Parameters:

trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)

set(ana_trig_filter: TriggerCoupling, trigger=Trigger.Default) None[source]
# TRIGger<*>:ANEDge:FILTer
driver.trigger.anEdge.filterPy.set(ana_trig_filter = enums.TriggerCoupling.LFReject, trigger = repcap.Trigger.Default)

Sets a filter for the external trigger signal to reject high or low frequencies.

Parameters:
  • ana_trig_filter

    OFF | LFReject | RFReject

    • OFF: The trigger signal is not filtered.

    • LFReject: Frequencies higher than the Cut-off frequency are rejected, lower frequencies pass the filter.You can adjust the Cut-off frequency using the TRIGgert:ANEDge:CUToff:LOWPass command, the default is 50 kHz.

    • RFReject: Frequencies below the Cut-off frequency are rejected, higher frequencies pass the filter.You can adjust the Cut-off frequency using the TRIGgert:ANEDge:CUToff:HIGHpass command, the default is 50 kHz.

  • trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)