FilterPy
SCPI Command :
TRIGger<*>:ANEDge:FILTer
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(trigger=Trigger.Default) TriggerCoupling[source]
# TRIGger<*>:ANEDge:FILTer value: enums.TriggerCoupling = driver.trigger.anEdge.filterPy.get(trigger = repcap.Trigger.Default)
Sets a filter for the external trigger signal to reject high or low frequencies.
- Parameters:
trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)
- set(ana_trig_filter: TriggerCoupling, trigger=Trigger.Default) None[source]
# TRIGger<*>:ANEDge:FILTer driver.trigger.anEdge.filterPy.set(ana_trig_filter = enums.TriggerCoupling.LFReject, trigger = repcap.Trigger.Default)
Sets a filter for the external trigger signal to reject high or low frequencies.
- Parameters:
ana_trig_filter –
OFF | LFReject | RFReject
OFF: The trigger signal is not filtered.
LFReject: Frequencies higher than the Cut-off frequency are rejected, lower frequencies pass the filter.You can adjust the Cut-off frequency using the TRIGgert:ANEDge:CUToff:LOWPass command, the default is 50 kHz.
RFReject: Frequencies below the Cut-off frequency are rejected, higher frequencies pass the filter.You can adjust the Cut-off frequency using the TRIGgert:ANEDge:CUToff:HIGHpass command, the default is 50 kHz.
trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)