Range
SCPI Command :
TRIGger<*>:SLEW:RANGe
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(trigger=Trigger.Default) SlewRangeMode[source]
# TRIGger<*>:SLEW:RANGe value: enums.SlewRangeMode = driver.trigger.slew.range.get(trigger = repcap.Trigger.Default)
Defines the time limit for the slew rate in relation to the upper or lower trigger level (see method
RsRtx.trigger.slew.rate.set()and methodRsRtx.trigger.slew.delta.set()) . The time measurement starts when the signal crosses the first trigger level - the upper or lower level depending on the selected slope - and stops when the signal crosses the second level.- Parameters:
trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)
- Returns:
range_mode: INSRange | OUTRange | LTHan | GTHan
INSRange: Triggers on pulses inside a given range. The range is defined by the slew rate ±delta.
OUTRange: Triggers on pulses outside a given range. The range is defined by the slew rate ±delta.
LTHan: Triggers on pulses shorter than the given slew rate.
GTHan: Triggers on pulses longer than the given slew rate.
- set(range_mode: SlewRangeMode, trigger=Trigger.Default) None[source]
# TRIGger<*>:SLEW:RANGe driver.trigger.slew.range.set(range_mode = enums.SlewRangeMode.GTHan, trigger = repcap.Trigger.Default)
Defines the time limit for the slew rate in relation to the upper or lower trigger level (see method
RsRtx.trigger.slew.rate.set()and methodRsRtx.trigger.slew.delta.set()) . The time measurement starts when the signal crosses the first trigger level - the upper or lower level depending on the selected slope - and stops when the signal crosses the second level.- Parameters:
range_mode –
INSRange | OUTRange | LTHan | GTHan
INSRange: Triggers on pulses inside a given range. The range is defined by the slew rate ±delta.
OUTRange: Triggers on pulses outside a given range. The range is defined by the slew rate ±delta.
LTHan: Triggers on pulses shorter than the given slew rate.
GTHan: Triggers on pulses longer than the given slew rate.
trigger – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)