Mode
SCPI Command :
REFLevel<*>:AUTO:MODE
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(refLevel=RefLevel.Default) RefLevelAutoMode[source]
# REFLevel<*>:AUTO:MODE value: enums.RefLevelAutoMode = driver.refLevel.auto.mode.get(refLevel = repcap.RefLevel.Default)
Defines the high and low signal levels from which the reference levels are derived. This setting is only available for automatic reference level mode (see method
RsRtx.refLevel.ldetection.set()) .- Parameters:
refLevel – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘RefLevel’)
- Returns:
mode: AUTO | PPRobability | MPRobability | ABSolutepeak | UPLM | UMLP | UALM | UMLA
AUTO: Auto select absolute probability: most suitable signal levels for the selected measurement
PPRobability: Peak probability: signal levels with the highest probability value
MPRobability: Mean probability: signal levels with mean probability
ABSolutepeak: Absolute peak: absolute peak signal levels
UPLM: Upper absolute peak, lower mean probability: high signal level is the upper absolute peak, low signal level is the level with the mean probability in the lower half of the histogram.
UMLP: Upper mean probability, lower absolute peak: high signal level is the level with mean probability in the upper half of the histogram, low signal level is the lower absolute peak.
UALM: Upper absolute peak, lower manual value: high signal level is the maximum result value of the amplitude measurement; low signal level is manually set using REFLevelrl:RELative:LOWer.
UMLA: Upper manual, lower absolute peak: The upper signal level is set manually using REFLevelrl:RELative:UPPer. The lower level is the minimum result value of the amplitude measurement.
- set(mode: RefLevelAutoMode, refLevel=RefLevel.Default) None[source]
# REFLevel<*>:AUTO:MODE driver.refLevel.auto.mode.set(mode = enums.RefLevelAutoMode.ABSolutepeak, refLevel = repcap.RefLevel.Default)
Defines the high and low signal levels from which the reference levels are derived. This setting is only available for automatic reference level mode (see method
RsRtx.refLevel.ldetection.set()) .- Parameters:
mode –
AUTO | PPRobability | MPRobability | ABSolutepeak | UPLM | UMLP | UALM | UMLA
AUTO: Auto select absolute probability: most suitable signal levels for the selected measurement
PPRobability: Peak probability: signal levels with the highest probability value
MPRobability: Mean probability: signal levels with mean probability
ABSolutepeak: Absolute peak: absolute peak signal levels
UPLM: Upper absolute peak, lower mean probability: high signal level is the upper absolute peak, low signal level is the level with the mean probability in the lower half of the histogram.
UMLP: Upper mean probability, lower absolute peak: high signal level is the level with mean probability in the upper half of the histogram, low signal level is the lower absolute peak.
UALM: Upper absolute peak, lower manual value: high signal level is the maximum result value of the amplitude measurement; low signal level is manually set using REFLevelrl:RELative:LOWer.
UMLA: Upper manual, lower absolute peak: The upper signal level is set manually using REFLevelrl:RELative:UPPer. The lower level is the minimum result value of the amplitude measurement.
refLevel – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘RefLevel’)