Range

SCPI Command :

SEARch:TRIGger:SLEWrate:RANGe
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(search: str) SlewRangeMode[source]
# SEARch:TRIGger:SLEWrate:RANGe
value: enums.SlewRangeMode = driver.search.trigger.slewrate.range.get(search = 'abc')

Selects how the time limit for the slew rate is defined. The time measurement starts when the signal crosses the first trigger level - the upper or lower level depending on the selected slope - and stops when the signal crosses the second level.

Parameters:

search – String with the name of the search

Returns:

range_mode: INSRange | OUTRange | LTHan | GTHan

  • INSRange: Triggers on slew rates inside a given time range. The range is defined by Slew rate and ±Delta.

  • OUTRange: Triggers on slew rates outside a given time range. The range definition is the same as for Within range.

  • LTHan: Triggers on slew rates shorter than the given Slew rate limit.

  • GTHan: Triggers on slew rates longer than the given Slew rate limit.

set(search: str, range_mode: SlewRangeMode) None[source]
# SEARch:TRIGger:SLEWrate:RANGe
driver.search.trigger.slewrate.range.set(search = 'abc', range_mode = enums.SlewRangeMode.GTHan)

Selects how the time limit for the slew rate is defined. The time measurement starts when the signal crosses the first trigger level - the upper or lower level depending on the selected slope - and stops when the signal crosses the second level.

Parameters:
  • search – String with the name of the search

  • range_mode

    INSRange | OUTRange | LTHan | GTHan

    • INSRange: Triggers on slew rates inside a given time range. The range is defined by Slew rate and ±Delta.

    • OUTRange: Triggers on slew rates outside a given time range. The range definition is the same as for Within range.

    • LTHan: Triggers on slew rates shorter than the given Slew rate limit.

    • GTHan: Triggers on slew rates longer than the given Slew rate limit.